Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26010721026010720750001

LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry

Symptom

Fast Cals Aborted See Long Description Scanner Hardware stopped Scan during Fast Cals

Error message

Error: 260107210 Gantry (Axial) Velocity OVERSPEED - Hardware Detection. Thu Dec 9 08:37:45 2004 Host: Table Gantry Processor Ermes# : 260107210 Exception Class: Safe State Severity: Primary File: amc_isr.c Line: 527 Function: AXIAL CONTROL : Axial Control Scan Type: None/Unknown Scan 0/0/0 Gantry (Axial) Velocity OVERSPEED - Hardware Detection. Measured Velocity: 0.999 sec/rev (Reference: 1000 AMC State: 8 AMC Cmd:21) Hdw Limit: 360 msec/rev HW Error: A RESET of the Scan Hardware is required and try again. Thu Dec 9 08:37:45 2004 Error: 260107210 Host: TGP Process: tgp_subsys File: amc_isr.c Line: 527 Running diagnostic: Axial2 Functional Thu Dec 9 08:37:45 2004 Host: Table Gantry Processor Ermes# : 260107210 Exception Class: Safe State Severity: Primary File: amc_isr.c Line: 527 Function: AXIAL CONTROL : Axial Control Scan Type: None/Unknown Scan 0/0/0 Gantry (Axial) Velocity OVERSPEED - Hardware Detection. Measured Velocity: 0.999 sec/rev (Reference: 1000 AMC State: 8 AMC Cmd:21) Hdw Limit: 360 msec/rev HW Error: A RESET of the Scan Hardware is required and try again.

Error: 260107207 Axial Motor Control (AMC) logic will NOT accept commands in Shutdown Mode of Operation. A RESE

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.