AMX2 & AMX110 - SN2658A replaced by SN3661
Symptom
Service note 2658A for AMX2 and AMX110: Old style battery chargers do not have the added redundant circuitry for battery overcharge protection that the latest style charger provides. Old style chargers may fail to terminate charging, resulting in overcharging and damage to the batteries.
Applies to
- Product
- Portables
- Component
- AMX110 AMX2 AMX3
- Area
- X-ray R&F
Full cause & step-by-step resolution