Def8000 If the tube warm-up function is invoked from the worklist and then exited (with or without completing…

Definium 8000 Wallstand Tracking After a Tube Warm-up May Cause Confusion

Symptom

Def8000 If the tube warm-up function is invoked from the worklist and then exited (with or without completing the warm-up), the positioner still is in a tube warm-up offset mode. If an operator then does an auto position to the wallstand, the OTS will align offset to the detector. Another manifestation is that the operator may reposition the wallstand detector but the OTS will track with an offset giving the impression that there is an alignment problem.

Applies to

Models
Definium
Product
Definium 8000
Component
Wallstand (Thunder)
Area
X-ray R&F

Full cause & step-by-step resolution