GE Healthcare · X-ray R&FDefinium
Def8000 The touchscreen calibration is designed to be done at 0 deg.
Definium 8000 OTS UIF Touchscreen Calibration Must Be Done At 0 Deg
Symptom
Def8000 The touchscreen calibration is designed to be done at 0 deg. The calibration workflow places the touchscreen calibration after a step where the UIF is at 90 degrees making it easy for the FE to attempt the calibration with the UIF in the wrong orientation.
Applies to
- Models
- Definium
- Product
- Definium 8000
- Component
- OTS
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- AEC Backup Time message
- Definium 8000 Auto positioning does not work after a reset/reboot due to wall stand receptor at a hard limit during rebooting or repowering.
- Definium 8000 One or both monitors display the following message on every boot
- Definium 8000 Wallstand or Table Only Configurations running VolumeRAD M3 Software.
- Definium 8000 OTS Collimator has no power due to a tripped breaker (CB8) ont he Tube Suspension Logic (TSL) Board.
- Definium 8000 The Siemens Collimator has not power due to a tripped breaker (CB9) on the Tube Suspension Logic (TSL) Board.