XQi, XRd, XRd2X, Def8000, AMX700 Auto Paste exams crash on Leg AP Anatomy-View.

Revolution XQi, XRd, XRd2X, Def8000, AMX700 Auto Paste crashes on Leg AP exams

Symptom

XQi, XRd, XRd2X, Def8000, AMX700 Auto Paste exams crash on Leg AP Anatomy-View. This is a known issue that occurs when the Service Engineer incorrectly Restores “IPLooks” from an older product onto the Definium 8000 system. The older product (e.g., XR/d-2x Lightning version) does not contain image processing looks for Auto Paste anatomies. When this image processing file is incorrectly loaded onto a Definium 8000 product, the Auto Paste gives a failure/crash during the exam. The problem is 100% reproducible and does not occur under normal, correct system operation.

Applies to

Models
Revolution, Definium
Product
Revolution XQi Definium 8000 Definium AMX 700 Revolution XRd2X Revolution XRd
Area
X-ray R&F

Full cause & step-by-step resolution