Pular para o conteúdo principal
GE Healthcare · MRSigna2218537222197622333962236698

Unable to Scan Diffusion Weighted EPI Due to Poor Seating of CERD Receiver # 1.

Symptom

Unable to Scan DWI

Error message

2218537 Scan Stopped, by Cerd Error Server Request Fri Jan 18 16:01:10 2002 Error: 2218537 Host: lx-mr Process: scn File: MsaScanMgr.m Line: 2487 Scan stopped, by Cerd Error Server request.

2221976 FAST RECEIVER DIAGS CONFIG ERROR 2221976 "FAST RECEIVER DIAGS CONFIG ERROR Fast Receiver diags were selected but a Fast Receiver could not be detected in the system. Please check the Fast Receiver connections if it is installed."

2233396 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan. Thu Feb 22 17:59:07 2007 Error: 2233396 Host: MENEMR Process: NSP File: TPS:cap: c_int45 Line: 4 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan.

2236698 CERD Diags Analog Receiver Channel Present LO Test Failure. Receiver channel 1 is missing LO1 or LO2 per RF Receiver Status register. Error: 2236698 File: diagsMain.m Line: 2039 CERD Diags : Analog Receiver Channel Present LO Test Failure Receiver channel 1 is missing LO1 or LO2 per RF Receiver Status register. Check receiver module installation. Error: 2236698 File: diagsMain.m Line: 2039 CERD Diags : Analog R

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.