Replaced bad I or Q heliax cable leading to the body coil
Symptom
Arcing Heliax cable One of the blue 1/2 wavelength I and Q Heliax cables connected from the body hybrid to the body coil was bad. When they disconnected it, the center pin stuck in the hybrid splitter. They found on examining it that it had been arcing.
Error message
Spurious DD/TR Driver Fault
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-328147G1 Run 743 RF Cable from Hybrid Splitter J3 to Body RF Coil J2 MG3 A3 A1 J3 to MG2 A12 A1 J2. 10 ft Coax w/M&F “N”. comments: "Body Coil Drive" 46-328147G2 Run 744 RF Cable from Hybrid Splitter J4 to Body RF Coil J1 MG3 A3 A1 J4 to MG2 A12 A1 J1. 10 ft Coax w/M&F “N”. comments: "Body Coil Drive"