XRd2X Dual-Energy bone images may appear to have a grid line or Moiré artifact.

Revolution XQi, XRd, XRd 2X (Lightning) Platform & Definium Systems Grid lines artifacts in Dual-Energy images

Symptom

XRd2X Dual-Energy bone images may appear to have a grid line or Moiré artifact. This occurs most commonly when the images are viewed on a Review monitor at less than 100% magnification. In other words, if the images are shrunk to less than 1:1 ratio. The problem is especially noticeable at 50% magnification on PACS. When the images are viewed minified, the grid lines can create a Moiré artifact. This often looks like a low-frequency pattern across the image. The artifact usually runs across the image (vertical bands).

Applies to

Models
Revolution, Definium
Product
Revolution XRd Definium AMX 700 Revolution XQi Definium 8000 Revolution XRd2X
Area
X-ray R&F

Full cause & step-by-step resolution