Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2600031255000226000312050042

Scanner hardware fault

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260003125 Did not detect a encoder index pulse when homing CAM A. See LD for more details Tue Oct 10 06:52:31 2006 Error: 260003125 Host: HSC Process: CCB Firmware File: aperture_home.c Line: 409 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50002/1/2 Type: Axial Exception Level: Pri/Soft Time: 06:52:31:395 Log Series: 42 Did not detect a encoder index pulse when homing CAM A.

Error: 260003120 An error occurred attempting to home CAM A or B See LD for complete error Mon Jan 10 09:04:41 2005 Error: 260003120 Host: HSC Process: CCB Firmware File: aperture_home.c Line: 193 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50042/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 09:04:41:350 Log Series: 57 An error occurred attempting to home CAM A. An error occurred attempting to home CAM B.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5404254 (replaces 2125453-3) Collimator cam motor driver amp - used on 1.X (QXI) 2.X, 3.X, 4.X and 5.X. Not used on VCT See LD for details 5404254 2125453-3 Collimator cam motor driver amp -used on 1.X (QXI) 2.X, 3.X, and 4.X. Not used on VCT Part #: 5404254 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Helios Collimator CAM Drive FRU: Yes Status: Orderable Part #: 2125453-3 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: HELIOS COLLIMATOR, CAM MOTOR DRIVER FRU: Yes Status: Orderable Comments: FLS - CAM MOTOR DRIVER HELIOS COLLIMATOR MS08992 9/3/98 M3 4TH QTR '98 HQT: VENDOR ORIENTAL MOTOR LEAD TIME WITH AIRFEIGHT 12 DAYS BY BOAT 6 WEEKS. INCREMENTAL COST FOR AIR FREIGHT $2
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.