Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCTHiSpeed21425050816115014214054115031

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Scanner Hardware Stopped Scan with High Voltage errors

Error message

Error: 214250 KV Fault Detected: Cathode Overvoltage See Long Description for full error Tue Mar 27 14:31:08 2007 Error: 214250 Host: OBC Process: Diag Exec File: isr1_handler.c 1.35 Line: 248 Function: Diagnostics XRAY GENERATION : KV Loop Test Name: XRay - Manual Unique ID: 50816 Iteration: 1 Exc. Level: Hard Time: 115014 Error Type: Hardware KV fault detected: Cathode overvoltage. KV address: FFCFF3H Bit: D1

Error: 214054 KV board HV status error The KV board failed to send a High Voltage off level 3 interrupt See LD for details Tue Mar 27 14:31:09 2007 Error: 214054 Host: OBC Process: Diag Exec File: technic_monitor.c 1.40 Line: 1104 Function: Diagnostics XRAY GENERATION : Rotating HVON Control Test Name: XRay - Manual Unique ID: 50816 Iteration: 1 Exc. Level: Soft Time: 115031 Error Type: Hardware KV board HV status error. The KV board failed to send a High Voltage off level 3 interrupt. Anode: 6.4KV Cathode: 0.0KV KV HV off trip point: 10KV HV status: Address: FFCFF9H Bit: D0 Value: 1

Applies to

Models
LightSpeed, VCT, HiSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2205390 CATHODE HV Tank used on 1.X, 2.x, 3.x, 4.x and HiSpeed LightSpeed Systems - Replaces 2161306 See Long Description for FEMC information and picture 2205390 CATHODE HV Tank used on 1.X, 2.x, 3.x, 4.x and HiSpeed LightSpeed Systems - Replaces 2161306 Part #: 2205390 Part Type: GP REPAIRABLES Part Weight: 30.06 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: BLOC HT CT FAST - POUR EPO FRU: Yes Comments: FLS - HV TANK CT FAST (-) CATHODE REPLACES 2161306 HDQ: 5 IN TRANSIT FROM GEMSE 4/3/98. REPLACES 2161306. PER G.ORME, ONLY DIFFERENCE IS PACKAGING, SO T3# REMAINS THE SAME. C.BRICKLER 4/6/98 ****NOTES FOR CD TEAM************************* ALL PARTS ON HAND HAVE BEEN REPAIRED USING THE NEW PROCEDURES. DO NOT USE THE 2205390H2 PART NUMBER ANYMORE.MS26561 DSM 11/16/01 ** FE ASSIST ** The HV adapter/Spacer part number 2171866 Kit that is included with the HV tanks But it is only used on older Highspeed Gantry due to the smaller bolt size. Refer to MI&CT ESERVICE NOTE SN07091114a for additional details. (212001261 21-Jan-2009) *** 2ND OPINION NOTES *** Check Oil in wells. Prior to ordering tube INSTALL NEW TUBE should be run to regenerate calseed values and set up the gencal table. The tube should be isolated and the KV loop test run to verify with 90% degree of certainty the tube is or is not at fault 4/7/98 2205390 1.0 0.5 SEC/REV CATHODE HV SUPPLY
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.