GE Healthcare · CTLightSpeedVCT260003142500012600031411315518414217231
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
Missed ISD See Long Description Inter Scan Delay ISD Missed .
Error message
Error: 260003142 The Filter failed to move to the requested position See long description for more details Sun Apr 8 06:41:17 2007 Error: 260003142 Host: HSC Process: CCB Firmware File: filter_move.c Line: 104 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50001/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:41:17:580 Log Series: 46 The Filter failed to move to the requested position. Error: 260003141 Filter homing problem. See the Long Description for the full error code Fri Mar 4 21:22:12 2005 Host : Collimator Ermes # : 260003141 Exception Class : Abort Severity : Primary File : filter_home.c Line# : 147 Function : No System Function Reported Scan Type : None/Unknown Scan: 13155/1/0 Filter homing problem. Error: 184142 ISD Missed See Long description for complete error Wed Nov 8 11:01:40 2006 Error: 184142 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1043 Function: No System Function Reported Scan: 496/2/2 Type: Axial Exception Level: Pri/Soft Time: 11:01:40:295 Log Series: 937 ISD missed. Rx ISD: 1000 mS Requested ISD = 1 Second Actual ISD: 17231 Actual ISD = 17 Seconds More Data Follows: See Error: 184188 for additional de
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5140001-3 (replaces 5140001-2) Helios 3 collimator with notched filler bar and rating plate (used on 1.X-4.X) See Long Description for more info Part #: 5140001-3 Part Type: GP REPAIRABLES Part Weight: 115.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 3 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: ** FE Assist ** 1.x 2.x 3.x 4.x started with Ultras added minor quality improvements to the dash 2 and is backwards compatible (12/6/06 212001261) Part #: 5140001-2 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 2 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: This part is replacement for 2214768-2, please do not use stock of 2214768-2. ms13452 050516 Can use the 5140001-3 if no Stock of the -2 (9/26/05 212001261)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required