GE Healthcare · MRSigna2213312

auto prescan fails with erbtec 85 fault, but only on sagittal T2 scan.

Symptom

Auto Prescan Fails Only on T2 Scan

Error message

2213312 ERBTEC 20KW RF Amplifier fault 85 RF Power Monitor Fault ERROR INFO: 14:15:47.....23-Sep-97.........................Error Code: 2213312 STAT 00 in [ ST MONITOR] Abcode: 0 /srv ERBTEC 20KW RF Amplifier fault 85. RF power monitor fault. The 12v externally controlled HV safety relay on the Processor Board was disabled by the GE RF power monitor. Possible defective GE power monitor, cables, or Processor Board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
8.3 8.3
Area
MR