Fails TPS Reset when booting up to Applications or when Resetting TPS.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2244604 The GP reported more than one SGA Power Supply Overtemperature Faults. Thu Apr 14 15:00:00 2005 Error: 2244604 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 1726 The GP reported more than one SGA Power Supply Overtemperature Faults. Refer to prior fault message. Power cycle ACGD POWER breaker on the PDU to recover from this condition.

2244585 The SGA Power Supply failed to go to Ready when commanded, or came out of Ready 2244585 "The SGA Power Supply failed to go to Ready when commanded, or came out of Ready. If ENABLE test point on the front SGA-PS J2 connector is HIGH, replace the SGA-PS. Else, check the GP to SGA-PS cable, or replace the GP."

2226629 Gradient Identification Problem The gradient subsystem could not identify itself as 8915 or 8905 gradients. 2226629 Gradient Identification Problem. The gradient subsystem could not identify itself as 8915 or 8905 gradients. Possible gradient amplifier problem, please run diagnostics.

2244589 The GP reported a SGA Power Supply OverTemp Fault. Thu Apr 14 14:59:38 2005 Error: 2244589 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 1709 The GP reported a SGA Power Supply OverTemp Fault. Wait at leas

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.