No Gantry motions on Ventri via the RCU

Symptom

No Gantry motions on Ventri via the RCU No Gantry motions on Ventri via the RCU. System reboot had no effect. Gantry reset would pass but system startup would fail. Checking the error logs revelaed the following -Communications error with Gantry Control Unit -Internal motion sw error -Last axis movement failed -Comm/sync error with {0} -Cannot perform the requested operation because gantry is not ready Diags on the GCU internal test failed the following -GCU voltage test -Battery Ram memory test -GCU interrupt test -FPGA Bus test Contributor 212004100 Frastaci March 20 2007

Error message

STSMANGER_GCU_COMM_FAILURE:.:Communication error with the Gantry Control Unit.:.:1. Perform full shutdown. 2. Perform Gantry reset. 3. Perform PC reboot.:.: Communication failure with the GCU board (no TCP/IP connection):.:Trace:{0}