GE Healthcare · MRSigna2227367
Unable to scan.
Symptom
Unable to scan. Have popup message "spmon thick out of range" The Customer is trying to do a scan and it fails to setup and gives a popup message: "spmon thick out of range".
Error message
spmon thick out of range popup error seen on the applications screen during the protocol setup 2227367 Use an offset =6.2mm. Thu Feb 19 17:30:25 2004 Error: 2227367 Host: bmr1 Process: scn File: PsdError.m Line: 329 Use an offset < 150mm or use a slice thickness >=6.2mm.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.0
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required