GE Healthcare · X-ray R&FDefinium
Def8000 System experienced intermittent failures with either automatically pushing images to PACS on closed ex…
Definium 8000 Intermittent issue pushing images to PACS
Symptom
Def8000 System experienced intermittent failures with either automatically pushing images to PACS on closed exam or on manual push. These images may actually succeed in going across to PACS on retries.
Applies to
- Models
- Definium
- Product
- Definium 8000
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- AEC Backup Time message
- Definium 8000 Auto positioning does not work after a reset/reboot due to wall stand receptor at a hard limit during rebooting or repowering.
- Definium 8000 One or both monitors display the following message on every boot
- Definium 8000 Wallstand or Table Only Configurations running VolumeRAD M3 Software.
- Definium 8000 OTS Collimator has no power due to a tripped breaker (CB8) ont he Tube Suspension Logic (TSL) Board.
- Definium 8000 The Siemens Collimator has not power due to a tripped breaker (CB9) on the Tube Suspension Logic (TSL) Board.