Can not Run SPT Body SNR subtest.

Symptom

SPT Body SNR subtest Prescan Fails with Gradient Error

Error message

2245177 GRADIENT DRIVER TESTS - Dynamic Test failed on __ Axis Thu Sep 16 14:06:38 2004 Error: 2245177 Host: MRS5 Process: diagCoord File: grad_lib.c Line: 416 Diagnostic: GDT Stat&Dyn Iteration: 1 of 1 Action: 1 Task ID: 0x3fc9990 Time: 43455 Object: SCP Test Task GRADIENT DRIVER TESTS - Dynamic Test failed on X Axis. IERROR did not settle to +/- 0.031A in 0.000usec on the 0 transition. Actual= 0.000usec. Check GRAM Tuning, SGA, SGA Power Supply, and GP.

2244592 SGA Power Supply Under Voltage Fault 2244592 "SGA Power Supply Under Voltage Fault.\n\ Look at SGA-PS front LEDs. If BUS_OV LED lit, check PDU\n\ 420V output. If UV_X,Y,Z LED lit, swap military connectors\n\ to isolate SGA versus SGA-PS."

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution