less image space than normal after removing all the image data.
Symptom
Less image space available than normal. See long description for examples. Less image space available than normal, even after removing all exams in Image Works. Example: H3 or GOC1 console should have 20,000 (512 images) available on a clean system with an Octane cpu with two 9 GB disk drives. But with the replacement of the 9GB drives with 18 and 73GB drives, image space may increase up to 130,000 (512 images) depending on software. Example: GOC2 console should have 100,000 (512 images) available on a clean system with an HP-8000 cpu with one 18GB disk and one 73GB disk. Example: GOC3 or 4 console should have 200,000 (512 images) available on a clean system with an HP-8200 cpu with one 36GB disk and two 73GB disks. Data above is approximate, but can be used to determine if there is a problem with disk space.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 2.4.2_H2.4M5 for LightSpeed 2.X (Plus, QX/I) with SGI Octane Consoles running the IRIX OS 2.4.2_H2.4M5 for LightSpeed Plus 2.x with SGI Octane Consoles running the IRIX OSPart # Desc.2395178 - LS2002 M5 SDAS Software Collector 2395182 - LS2002 M5 Application Collector 2395160 - Apps Disk 1 2395161 - Apps Disk 2 2356012 - Irix OS Collector 2356015 - IRIX OS Disk 1 2356016 - IRIX OS Disk 2 2356017 - IRIX OS Disk 35113667 - LS2002 M5 Patch C.12395172 - LightSpeed Class C - 1104I.2_H4.0M5CLASSC2395173 - LightSpeed Class M - 1204I.2_H4.0M5CLASSM 2.4.2_H2.4M5 for LightSpeed Plus 2.x with SGI Octane Consoles running the IRIX OS FMI 25332 - LS2002 M5 Software Upgrade The purpose of this FMI is to upgrade LightSpeed IRIX-based CT scanners to the latest version software. This software includes safety and quality fixes. Safety (PSR / FPR) fixes include: • PSR 1286436 -- Auto mA table values, • PSR 1286629 -- Auto mA noise index, • PSR 13000489 -- Scan abort with prep delay, and • PSR 1286829 -- Wrong Patient Information. Quality fixes include: • Scout not displayed, • ring artifact, • remote tilt, • align light, • Autolink page slow, • Batch Install Image not displayed, • Image out of order, • Slow image-biopsy, • Transfer to Pacs Error, • Service Desktop Error, • autoMA-concerns (table), • Modem, • autoMA-noise (GUI), • death of stclf, • axial not ready, • Smart Prep fails to start, • .rhosts file corrupted, • scan data missing in retro, • Square VP not saved, • AutoArchive, • peripheral gate hang, • ali pacs (mekesson), • Can't recover MOD images, • not accept image - Phillips, • Show Local lines delay, and • Head IQ. New delivered features are: • AutomA GUI, • Repeat Series, • Re-Start / Shutdown Procedure, and • Anonymous Scan Data Save. Furnished Materials Qty Part number Description 1 2399103-100 FMI document 1 2395178 LS2002 M5 SDAS Software Collector containing: 1 2395182 LS2002 M5 Application CD Collector (3 CDs) 1 2356012 LightSpeed OS Software Collector (3 CDs) 1 46-302200P2 Product Locator 1 2348269-100 LS2002 Software Install Document 1 2243321-100 LightSpeed 2X Apps Tips and Workarounds 1 2369740-200 CT Lightspeed Series Learning and Reference Guide Tools Required Qty Part # Description 1 2395172 LightSpeed Class C Software (Tab Distribution) 1 2395173 LightSpeed Class M Software (Tab Distribution) 1 2347142 Variviewer Software Option (Customer Site) 1 2356009 Auto mA Software Option (Customer Site) 1 2345411 Power 440 Software Option (Customer Site) - - Customer Purchased Options (Customer Site) Synonyms: LSPlus, B7850JE, B7850JF, B7850JL Gantry: 2260619, 2260619-3, B7850NG Console: 2266832 2266832-2 2304732, 2304732-2, B7850CD B7850NC B7850ND
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required