GE Healthcare · MRSigna2225643

Erbtec Faults Due to Door Knob Capacitor Arcing In RF Amplifier.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2225643 ERBTEC RF Amplifier fault 57 +3kV power supply unacceptable within 0.5 to 6 seconds after closing HV start relay 2225643 ERBTEC RF Amplifier fault 57. +3kV power supply unacceptable within 0.5 to 6 seconds after closing HV start relay. Possible defective HV rectifier/filter board, bad relay in AC Switching Module, or bad A/D on Processor Board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
8.3 8.3
Area
MR