GE Healthcare · MRSigna2225643
Erbtec Faults Due to Door Knob Capacitor Arcing In RF Amplifier.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2225643 ERBTEC RF Amplifier fault 57 +3kV power supply unacceptable within 0.5 to 6 seconds after closing HV start relay 2225643 ERBTEC RF Amplifier fault 57. +3kV power supply unacceptable within 0.5 to 6 seconds after closing HV start relay. Possible defective HV rectifier/filter board, bad relay in AC Switching Module, or bad A/D on Processor Board.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 8.3 8.3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required