PROTEUS JEDI 50-0204 Filament Error

Symptom

Error 50-0204 Filament permanent open circuit Intermittent Error Code 50-0204 (Filament permanent open circuit).

Error message

50-0204 Jedi error code See long description for details Filament permanent open circuit - Issued by the Heater board when an open has been detected and 3 restarts have been tried without success within 100 ms. Potential causes: 1. X-ray tube filament open 2. Heater to HV Tank cable 3. Cathode HV cable or pin contacts 4. Open circuit in filament transformer inside HV tank.

Applies to

Product
PROTEUS XRa
Component
Rad 21 X-ray tube JEDI Generator This generator is used on Proteus and Precision 500D systems.
Area
X-ray R&F