GE Healthcare · MRSigna2000023682000023522223639
Cannot Boot Applications with Host Monitor Timeout Error
Corrupt Software.
Symptom
Cannot Boot Applications with Host Monitor Timeout Error
Error message
Host Monitor Timeout after 200 seconds 200002368 A set of processes are deadlocked waiting for each other to start. Mon Apr 12 11:43:42 2004 Error: 200002368< Host: BaptistGEMR Process: hostMonitorbsp File: ssProcStart.c Line: 1099 A set of processes are deadlocked waiting for each other to start. svat, scn, rtcProxy, resetServer, lais, hgp, displayServer, autoview, app, WLServer, TIR, SITT, RtpHost, RTPSIServer, RIR, PPSproxy are the likely culprits. This deadlock has caused the system to shutdown! Failed to get machine name from NSP Fri Jan 28 10:36:31 2005 Error: 0 Host: lx-mr Process: checkForSecurit yKey File: mrmailmgr.c Line: 222 Failed to get machine name from NSP Hostmonitor timeout after 150 seconds 200002352 A HOST_SHUTDOWN was initiated by the death of RTPSIServer See Long Description 200002352 Host: lx-mr Process: hostMonitor File: ssProcStop.c Line: 810 A HOST_SHUTDOWN was initiated by the death of RTPSIServer. 2223639 Environmental variable TPSHOST is not defined Thu Oct 23 18:03:37 2003 Error: 2223639 Host: HDC_MRI1 Process: checkForSecurityKey File: mrmailmgr.c Line: 240 Environmental variable TPSHOST is not defined.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.