Drive Issues with portable unit

Symptom

Intermittent drive issues Some newer AMX4+ units have been encountered that make a scuffing noise when driven, especially at lower speeds and particularly when going around a corner. The problem lies with the drive motor mount bolts being tightened down too much from new, preventing belt tensioning springs from normal operation. After usage for a few months the belts stretch slightly, and the noise is actually the drive belt slipping on the pulleys.

Applies to

Product
Portables AMX4+
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.