System configured as a Hispeed vs. an Echospeed.

Symptom

BACC: FE unable to complete any of the calibrations for the Fast Receiver All of the calibrations where the fast receiver value was set to zero worked fine. Fast Receiver installed physically on the system.

Error message

2233396 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan. Thu Feb 22 17:59:07 2007 Error: 2233396 Host: MENEMR Process: NSP File: TPS:cap: c_int45 Line: 4 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan.

2234205 The RAP timed out during acquisition. Internal software or hardware error, please run CERD diagnostics! Thu Feb 22 17:59:07 2007 Error: 2234205 Host: MENEMR Process: NSP File: TPS:asc_ba_isr Line: 1 The RAP timed out during acquisition. Internal software or hardware error, please run CERD diagnostics!

2233420 Data from the fast receiver has overflowed the data acq. task Please re-try scan at lower bandwidth! Thu Feb 22 17:59:17 2007 Error: 2233420 Host: MENEMR Process: NSP File: TPS:cap: ca_epi_int.c Line: 146 Data from the fast receiver has overflowed the data acq. task Please re-try scan at lower bandwidth!

2218537 Scan Stopped, by Cerd Error Server Request Fri Jan 18 16:01:10 2002 Error: 2218537 Host: lx-mr Process: scn File: MsaScanMgr.m Line: 2487 Scan stopped, by Cerd Error Server request.

Applies to

Models
Signa, HiSpeed
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR