GE Healthcare · MRSignaHiSpeed2233396223420522334202218537
System configured as a Hispeed vs. an Echospeed.
Symptom
BACC: FE unable to complete any of the calibrations for the Fast Receiver All of the calibrations where the fast receiver value was set to zero worked fine. Fast Receiver installed physically on the system.
Error message
2233396 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan. Thu Feb 22 17:59:07 2007 Error: 2233396 Host: MENEMR Process: NSP File: TPS:cap: c_int45 Line: 4 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan. 2234205 The RAP timed out during acquisition. Internal software or hardware error, please run CERD diagnostics! Thu Feb 22 17:59:07 2007 Error: 2234205 Host: MENEMR Process: NSP File: TPS:asc_ba_isr Line: 1 The RAP timed out during acquisition. Internal software or hardware error, please run CERD diagnostics! 2233420 Data from the fast receiver has overflowed the data acq. task Please re-try scan at lower bandwidth! Thu Feb 22 17:59:17 2007 Error: 2233420 Host: MENEMR Process: NSP File: TPS:cap: ca_epi_int.c Line: 146 Data from the fast receiver has overflowed the data acq. task Please re-try scan at lower bandwidth! 2218537 Scan Stopped, by Cerd Error Server Request Fri Jan 18 16:01:10 2002 Error: 2218537 Host: lx-mr Process: scn File: MsaScanMgr.m Line: 2487 Scan stopped, by Cerd Error Server request.
Applies to
- Models
- Signa, HiSpeed
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.