Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2600073871211626000793950444260007323260007774

LightSpeed 5.X (80/100kW)

Symptom

Intermittent DAS communication errors. Problem occurs if scanning is paused or aborted. See Long Description Once fault occurs the exam needs to be ended and a scan hardware reset performed at a minimum to resume scanning again.

Error message

Error: 260007387 DCB hardware has detected a Parity Error. See Long Description for complete error Wed Feb 7 16:50:55 2007 Error: 260007387 Host: orp Process: orp_subsys File: parity_16.c Line: 271 Wed Feb 7 16:50:54 2007 Host : Das Ermes # : 260007387 Exception Class : Abort Severity : Primary File : parity_16.c Line# : 271 Function : No System Function Reported Scan Type : None/Unknown Scan: 12116/2/0 SCAN ABORT -- DAS Has Detected An Error. DCB hardware has detected a Parity Error. DCB firmware is UNABLE to determine the location of the Parity Error. Hardware Indicates that the ERROR is on (CardStream Index: 6, RowStream Mask: 0x0).

Error: 260007939 DAS Converter Board(s) encountered a divide by zero during alpha calibration. See Following Error Message for Converter Card List See long description for full error Tue Dec 28 09:41:31 2004 Error: 260007939 Host: orp Process: orp_subsys File: brdflt.c Line: 82 Tue Dec 28 09:41:31 2004 Host : Das Ermes # : 260007939 Exception Class : Soft Severity : Pri/Soft File : brdflt.c Line# : 82 Function : No System Function Reported Scan Type : None/Unknown Scan: 50444/1/0 DAS Converter Board(s) encountered a divide by zero during alpha calib

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues