Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260115244750001703000101663260111023

LightSpeed 5.X (60kW)

Symptom

Table will not move up, especially when in full down position. See long description Table will not raise or move up, especially when it's in the full down position and this could be intermittent along with noises noticed by the customer.

Error message

Error: 260115244 MTCB cradle moving direction error See the Long Description for the complete error code Thu Dec 28 10:59:18 2006 Error: 260115244 Host: tgp Process: tgp subsys File: mtcb cradle monitor.c Line: 823 Thu Dec 28 10:59:18 2006 Host : Table Sub-System Ermes # : 260115244 Exception Class : Secondary Severity : Pri/Soft File : mtcb cradle monitor.c Line# : 823 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 MTCB cradle moving direction error. Possible Encoder, cabling or MTCB board issue. Current Speed = -7300 Target Speed = 75000 Target Position = 1703000 Current Position = 101663

Error: 260111023 The pushbutton hardline did not clear after buttons were released. See long description for complete error Wed Oct 18 12:16:05 2006 Error: 260111023 Host: tgp Process: tgp_subsys File: guif_can_mgr2.c Line: 3270 Wed Oct 18 12:16:05 2006 Host : Table Gantry Processor Ermes # : 260111023 Exception Class : Secondary Severity : Pri/Soft File : guif_can_mgr2.c Line# : 3270 Function : OPERATOR INPUT/OUTPUT : Gantry UIF Communications Scan Type : None/Unknown Scan: 5621/2/0 The pushbutton hardline did not clear after buttons were released.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2337217 elevation encoder assembly used on HPower Systems. See Long Description for details Part #: 2337217 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ENCODER ASSY, ELEVATION - H POWER FRU: Yes Status: Orderable Parts Offered as: Current: 2337217 Harvested: Not Available Exchange: Not Available Comments: FLS - ENCODER ASSY, ELEVATION - H POWER. ADDED PER CT PCN 220839. MS20544 3/31/03.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.