Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260107205

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Axial motor drive fault. Gantry will not rotate

Error message

Error: 260107205 Axial Motor Control (AMC) logic could NOT clear the FAULT line from the AXIAL Drive See Long Description Tue Jun 6 16:50:06 2006 Error: 260107205 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 2605 Tue Jun 6 16:50:07 2006 Host : Table Gantry Processor Ermes # : 260107205 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 2605 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic could NOT clear the FAULT line from the AXIAL Drive. Last Axial Drive Fault will be logged before this message.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5118577-2 Axial Dynamic Braking Module ( Replaces 5118577 ) See long description for details Part #: 5118577-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Axial Dynamic Braking Module FRU: Yes Status: Orderable Comments: added per ECR, ms13452 051201 This part is replacement for 5118577, do not continue to use 5118577.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues