Pular para o conteúdo principal
GE Healthcare · MRSigna10097

Images not showing up in browser due to corrupt image data base.

Symptom

Images Reconstructed but not Listed in the Browser, Only Displayed in Auto viewer. See Long Desc. Caller had images in Auto Viewer but not in Browser. Caller had scanned several series since but had turned off Autoview prior to scanning the additional new series. When Autoview is turned off, it will hold onto the last images that were scanned with it turned on.

Error message

10097 Save Exam/Series Request Points to Empty Directories. 10097 Host: aicmr Process: arsjob File: SendErrorResult.c Line: 87 Save Exam/Series request points to empty directories.

1031 Stored Procedure Failed. Identifyimage failed Error: 1031 Host: aicmr Process: dbwserver File: batch.ec Line: 917 stored procedure failed : -211 - Identifyimage failed

9065 Internal Error, Fork not Allowed NO FREE PID 9065 Host: aicmr Process: dbwserver File: proc.c Line: 187 Internal error, fork not allowed NO FREE PID

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.0
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required