Grinding noise coming from the table while driving cradle in or out. Table may move slower than normal. Customer may hear a grinding noise coming from the table while driving cradle in or out. Table may move slower than normal.

Error message

Error: 182211 Velocity error detected by hardware (LM628 excessive position error). See the Long Description for the full error Tue Aug 29 08:09:30 2006 Error: 182211 Host: ETC Process: Cradle File: cmcLm628IsrTsk.c 1.19 10/16/9 Line: 185 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 20525/2/1 Type: Helical Exception Level: Pri/Soft Time: 08:09:30:260 Log Series: 595 HW error: Velocity error detected by hardware (LM628 excessive position error). Expected Velocity 75000 (.001 mm/sec) Actual Velocity 0 (.01 mm/sec) Pos Err Tol 2540 (.01 mm) Cur Pos 90852 (.01 mm) Cradle Dir: 1 (0 = Out 1 = In [Gantry]).

Error: 184862 Cradle unable to initiate a profile. Desired position 3880, velocity 47 internal cradle status 15006 See Long Description for full error Error: 184862 Host: ETC Process: Cradle File: cmc_interface.c 1.30 9/25/96 Line: 715 Function: No System Function Reported Scan: 8637/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 22:05:59:490 Log Series: 104 Cradle unable to initiate a profile. Desired position 3880, velocity 47 internal cradle status 15006

Error: 182213 Velocity tracking error See Long Description for the full error code Tue Mar 15 15:28

Applies to

Models
HiSpeed, LightSpeed
Product
HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues