GE Healthcare · MRSigna22257432225744
Will Not Scan With RF Errors.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2225743 Analogic RF Amplifier 1 Peak Overloaded Fault. Instantaneous peak drain current of RF devices exceeds 120 amps. Wed Dec 6 18:11:01 2000 Error: 2225743 Host: Mobile Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3066 Analogic RF Amplifier 1 I Peak Overloaded Fault. Instantaneous peak drain current of RF devices exceeds 120 amps.> 2225744 Analogic RF Amplifier 1 Device Fault 44. RF FET device is shorted. Replace amplifier. Analogic RF Amplifier 1 Device Fault 44 indicates the first of the two Analogic RF Amplifiers, the one that is mounted above the second Analogic RF Amplifier. 2225744 Analogic RF Amplifier 1 Device Fault 44. RF FET device is shorted. Replace amplifier.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR