Will Not Scan With RF Errors.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2225743 Analogic RF Amplifier 1 Peak Overloaded Fault. Instantaneous peak drain current of RF devices exceeds 120 amps. Wed Dec 6 18:11:01 2000 Error: 2225743 Host: Mobile Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3066 Analogic RF Amplifier 1 I Peak Overloaded Fault. Instantaneous peak drain current of RF devices exceeds 120 amps.>

2225744 Analogic RF Amplifier 1 Device Fault 44. RF FET device is shorted. Replace amplifier. Analogic RF Amplifier 1 Device Fault 44 indicates the first of the two Analogic RF Amplifiers, the one that is mounted above the second Analogic RF Amplifier. 2225744 Analogic RF Amplifier 1 Device Fault 44. RF FET device is shorted. Replace amplifier.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.