GE Healthcare · CTLightSpeedVCT290010004204632862204633118
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Cannot perform any scanning. Cannot perform any scanning
Error message
Error: 290010004 Checksum mismatch for view 1241. Expected 204632862 : Got See Long Description Tue Dec 6 11:07:48 2005 Error: 290010004 Host: darc Process: data_restore File: mdas_intf_object.cxx Line: 372 Checksum mismatch for view 1241. Expected 204632862 : Got 204633118
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Software
- 06MW03.4 For LightSpeed 7.X VCT - issued by FMI 25373 released 1st Quarter 2006. Reboot BEFORE and AFTER detailed cals 06MW03.4 For LightSpeed 7.X VCT - issued by FMI 25373 released 1st Quarter 2006. Reboot BEFORE and AFTER detailed cals 06MW03.4 For LightSpeed 7.X VCT - issued by FMI 25373 Titled VCT M4 (06MW03.4) Software Update. Released 1st Quarter 2006. Started shipping forward production on February 20th, 2006. Reboot BEFORE and AFTER detailed cals ClassC = part# 5167665 Ver 06MW03C.4 Tab distributed ClassM = part# 5167597 Ver 06MW03M.4 Tab distributed FMI 25373 Information: Effectivity The following CT Equipment is affected: All VCT systems manufactured before March 2006 Model number tracked: 5124069 Purpose The purpose of this FMI is to update the VCT installed base with VCT M4 software, documentation and to verify the correct wiring on the GFI breaker. Furnished Materials FMI 25373 Collector 5145027 Qty Description Part # 1 FMI 25373 Document 5144791-100 1 VCT Hi-power opt SW Key 5169414 1 Documentation Collector M4 5136582-3 1 VCT M4 06MW03.4 software collector 5167622 1 Split Ferrite Core 2338755 See FMI document for complete information
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.