88S/85SFD 8835/8535SFD Service Note 3139 Hall Effect Power Assist Handles

Symptom

Two errors were found in the information in the addenda after distribution of Service Note 3139. These errors were as follows: The jumper tables for JP1, JP2, JP3 and JP4 were in error in the instructions in Chapter 6 and on the 46-264316G1 schematic in Chapter 7 on both addenda. (All jumper positions were specified backwards.) (The gain adjustment statements should not have referenced adjusting for pounds of force.)

Applies to

Component
8535SFD 8835SFD
Area
X-ray R&F