Definium 8000 & Revolution XR/d 2X (Lightning) Replacement Table I/O Board 2290188

Symptom

Def8000 and XRd2X Replacement Table I/O board 2290188 appears to be DOA also RED light on motion control amp Two sites (a Lightning and a Thunder) had a suspected problem with the Table I/O Board. After replacement board was installed, the Motion Amp always had the status LED lit red. When the cable to P4 on the Motion Amp was disconnected, the status LED was lit green. Multiple replacements of the Table I/O Board, the Motion Amp, and the cable did not change the symptoms.

Applies to

Models
Definium, Revolution, Precision
Product
Definium 8000 Revolution XRd2X
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.