Legacy breaker trips intermittently when the table is angulating due to a defectvie RFP1 A8 CB1 breaker.

Symptom

Legacy breaker trips intermittently when the table is angulating. This breaker goes off intermittently when the table is tilting. The first Legacy units have a 10 amp breaker. When tilting the table, the current demand across this breaker reaches 12 amps, so the breaker trips.

Applies to

Component
Legacy Table
Area
X-ray R&F