Legacy No table or wall bucky exposure.

Symptom

Legacy No table or wall bucky exposure. No error codes. The real time bus cable not fully seated in the positioner cabinet. Legacy new install. No exposure to table or wall bucky. No error codes. Get prep, grid oscilates when the exposure switch is depressed. Wall bucky or table bucky ready signal is at the A9 interfiace card just before it goes into the bus line. Ultranet collimator has a green light and collimated to the correct size. Same problem with AEC or Non AEc exposures. Ran diagnostics Realtime bus tests. Test failed.

Applies to

Product
Table
Component
Legacy Table
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.