GE Healthcare · MRSigna
When they went to filming the left click wouldn't work, but the mouse cursor would move Due to PPS option had IP address…
When they went to filming the left click wouldn't work, but the mouse cursor would move Due to PPS option had IP address assigned but wasn't setup on broker.
Symptom
When they went to filming the left click wouldn't work , but the mouse cursor would move When they went to filming the left click wouldn't work , but the mouse cursor would move.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- M1033PA Perform Procedure Step Option M1033PA is "Perform Procedure Step Option", an optional enhancement to ConnectPro. The option key name in software shows up as "pps".
- Area
- MR
Full cause & step-by-step resolution