PROTEUS JEDI Filament Error 50-0204

Symptom

PROTEUS In addtion to the X-ray tube, not in auto mode, check if one of the filiments are broken by running the heater test. While looking at the heater board, ran the test and found that the LF was good and the SF caused errors.

PROTEUS JEDI Filament Error 50-0204 Customer reports intermittent Error 50-0204 (Filament permanent open circuit).

Error message

Error 50-0204 Jedi Filament

Applies to

Product
PROTEUS XRa
Component
JEDI Generator This generator is used on Proteus and Precision 500D systems.
Area
X-ray R&F

Full cause & step-by-step resolution