PROTEUS OTS Angulation Lock Diode There is a diode for the angulation lock that is located in the goose neck of the hanger. If this diode is shorted it will take out the 24V for the locks on the hanger and the locks won't work at all. F4 on the XT Control Board opens. This is a 3 AMP, 250V slow blow fuse.