Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1841814167426000682318729

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Gantry rotates but system will not scan. Gantry rotates but system will not scan

Error message

Error: 184181 Number of offsets acquired does not match estimate of number of offsets generated See Long Description for complete error Wed Nov 16 10:17:50 2005 Error: 184181 Suite: CT27 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.68 Method: No Method Line: 480 Function: XRAY GENERATION : Scan Data Trigger Acquisition Scan: 41674/201/1 Type: Axial Exception Level: Pri/Most Severe Time: 10:17:49:830 Log Series: 755 Number of offsets acquired does not match estimate of number of offsets generated HW ERROR: Possible hardware problem. Offsets generated (estimate): -4904 Difference Between Generated and Acquired: 9810 Test Margin: 12 Triggers

Error: 260006823 Gantry (Axial) could NOT find the ''C'' pulse. See Long Description for full error Tue Jan 17 16:51:34 2006 Error: 260006823 Host: STC Process: Axial File: amc_exec.c Line: 3248 Function: AXIAL CONTROL : axial control Scan: 18729/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 16:51:34:165 Log Series: 144 Gantry (Axial) could NOT find the ''C'' pulse. (AMC State:4 Cmd:0x150c0008 Cmd Vel: 0.008 sec/rev). HW/SW Error: Retry Rx to verify probable Gantry Hardware problem.

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.