AMX 4 - Unit moves very slowly in drive mode.

Symptom

AMX 4 - Unit moves very slowly in drive mode. The FE has replaced the Drive Control Board and the Drive Power Amp Board. The unit tries to drive, but basically does not drive, it seems like the brakes are on. When the manual brake button is utilized, the brakes release and the unit drives at the proper speed. Reference page 3 of the Drive Power Amp Board: The K145 relay is coming in when the handle is twisted. The only remaining difference is the relay uses +BAT SWITCHED signal to release the brakes.

Applies to

Product
AMX4+ Portables AMX4
Area
X-ray R&F