OrthoMRI Windows based system timeout error stops transmission of images to the remote host.

Symptom

OrthoMRI Timeout Error Stops Transmission of Images to the Remote Host The system is able to talk with the network and open the "AE_GW4" protocol but for some reasons the system has to wait for more than three minutes and the timeout stops the transmission. The system is able to talk with the network and open the "AE_GW4" protocol but for some reason the system has to wait for more than three minutes and the timeout stops the transmission.

Error message

OrthoMRI E TCP init Error Timeout see long description The following error, presents in the rec files, comes from the Merge library (DICOM) and not from our software: (1144) 07-25 10:21:22.17 MC3 T9: PDU Type: 04 (PDATA-TF) Sent (1144) 07-25 10:21:22.17 MC3 T9: Reserved Byte: 00 (1144) 07-25 10:21:22.17 MC3 T9: PDU length: 0000 4000 (16384 bytes) (1144) 07-25 10:21:22.17 MC3 T9: PDV length: 0000 3ffc (16380 bytes) (1144) 07-25 10:21:22.17 MC3 T9: Presentation Context ID: 01 (1144) 07-25 10:21:22.17 MC3 T9: Message Control Header: 00 (1144) 07-25 10:24:22.17 MC3(Send_PDV_List) E: TCP init error, timeout in TL_Write() (1144) 07-25 10:24:22.17MC3(Transport_Conn_Closed_Event)E:Transport unexpectedly closed

Applies to

Product
Orthopedics E-Scan (Win) XQ. Orthopedics E-Scan (Win) XQ.
Software
OrthoMRI 9.1
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.