GE Healthcare · CTBrightSpeedHiSpeed26011814426011801150075260118140
BrightSpeed 4/8/16 slice
Symptom
Unable to scan or perform tube warmup When the tube rotor spins up or down it may sound like a jet plane is landing and the gantry covers vibrate
Error message
Error: 260118144 (25-0532)LVPS400:heater0 + 160V too low. See Long Description for complete error. Mon Oct 23 08:02:17 2006 Error: 260118144 Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Mon Oct 23 08:02:16 2006 Host : Jedi Ermes # : 260118144 Exception Class : Secondary Severity : Sec/Soft File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Helical Scan: 8818/2/1 (25-0532)LVPS400:heater0 + 160V too low. ^C Z6,Z7= 82h,0h Error: 260118011 CAN_WRITE_OVERRUN. This error occurs when the library waits more than CAN_DATA_MUTEX_TIMEOUT to be able to send a message See LD for complete error Fri Feb 3 07:59:35 2006 Error: 260118011 Host: orp Process: orp_subsys File: gen_exec.c Line: 459 Fri Feb 3 07:59:34 2006 Host : On-Board Rotating Processor Ermes # : 260118011 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 459 Function : SYSTEM COMMUNICATIONS : Generator CAN Serial Link Scan Type : None/Unknown Scan: 50075/6/0 CAN_WRITE_OVERRUN. This error occurs when the library waits more than CAN_DATA_MUTEX_TIMEOUT to be able to send a message. Error: 260118140 (25-0528)LVPS400:can + 15V too low. See Long Description for complete erro
Applies to
- Models
- BrightSpeed, HiSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 99185290 FUSE FF 14X51 15A 700VDC C See Long Description for Details. Part #: 99185290 Description: FUSE FF 14X51 15A 700VDC C 50KA ULR Part #: 99185290 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUS FF 14X51 15A 700VDC C 50KA ULR FRU: Yes Status: Orderable Comments: Not Available 2409042 (new pn) 2325360-6 (old pn) LVPS 400 (250V) used in JEDI Aux Box of all Xi HiSpeeds and 5.X RT, 60kW 100kW LightSpeeds, Brightspeed 4/8/16 See LD for FEMC info and picture 2409042 (new pn) 2325360-6 (old pn) LVPS 400 (250V) used in JEDI Aux Box of all Xi HiSpeeds and 5.X RT, 60kW 100kW LightSpeeds picture below Note: This Item is listed as a FRU YES for the HP60 but even though the part number is the same, it is listed as a FRU NO for HP100 and RT. Part #: 2325360-6 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: LVPS 400 (250V) BOOT 2.05/APPLI 2.36 FRU: Yes Status: Orderable Comments: FLS - LVPS 400 (250V) BOOT USED ON: GENERATOR, JEDI 100 2.05/APPLI 2.36 VASC 1T ADDED PER SPCN: SUPPLIED BY GEMS INDIA 030903RF141542N1 MS13452 030908 RISK: SITE DOWN FREQUENCY: MEDIUM PER SPCN THIS PART SERVES AS REPLACEMENT FOR 2325360-5, ALL REMAINING STOCK OF -5 IS TO BE UPGRADED TO -6 VERSION. The only difference between pn 2409042 (new pn) and pn 2325360-6 (old pn) is packaging improvment . You can order pn 2325360-6 if no pole stock on pn 2409042 (6/2005 P.Liebau)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.