GE Healthcare · X-ray R&F
MPX R&F system, Intermittent no READY light after fluoro exposure
Symptom
MPX R&F system, Intermittent no READY light after fluoro exposure MPX R&F system, intermittent no READY light. When the system is turned on MPX Console READY light is ON. Exposures are allowed in RAD and Spot film mode. After a fluoro exposure is initiated and a spot film is taken, no rad or spot film exposures are allowed. READY light goes off due to loss of filament current. Relay MG3A4K19 REC to FL TRANS FIL CKT INTR is energized. If you tap on the relay, the relay drops out and READY light turns back on.