GE Healthcare · X-ray R&F
MPX Relay MG3A2K15 in the SARC cube turns off and on at about 3 times/second after power on.
Symptom
MPX Relay MG3A2K15 in the SARC cube turns off and on at about 3 times/second after power on. This relay turns off if an overcurrent occurs in the primary of the T1 transformer in the SARC cube MG3A2. If there is a short in the T1 transformer of the SARC, this relay de-energizes and opens the supply to the center tap of the T1 transformer. The current drops to zero and the relay energizes again. It will cycle through this ON/OFF sequence as long as there is an overcurrent condition in the transformer.
Applies to
- Product
- MPX
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required