MPX The safety contactor does not energize. -15V power supply missing

Symptom

MPX The safety contactor does not energize. -15V power supply missing Found -15v on the MG2A5A1 missing. Traced problem to a shorted C58 on the MG2A6A41 board. Replaced cap and system is working.