Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260116086

LightSpeed 5.X (60kW)

Symptom

When pressing Alignment Light Button, Gantry rotates but no lasers come on and the ORP reset indicator flashes on the gantry display

Error message

Error: 260116086 Scanning Hardware Reset is needed to recover See Long description for complete error Tue Oct 31 14:19:54 2006 Error: 260116086 Host: tgp Process: tgp_subsys File: sys_watchdog_err_handle.c Line: 84 Tue Oct 31 14:20:07 2006 Host : Table Gantry Processor Ermes # : 260116086 Exception Class : Safe State Severity : Primary File : sys_watchdog_err_handle.c Line# : 84 Function : No System Function Reported Scan Type : None/Unknown Scan: 9328/2/0 Application ID 211 has failed to report to the system watchdog appId refresh time = 3000 wdog tmo time = 100 Scanning Hardware Reset is needed to recover

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2105617-5 (replace 2105617-4) SINGLE, TERMINATED LASER DIODE ASM - used on H2 and HP gantry See long description for details SINGLE, TERMINATED LASER DIODE ASSEMBLY Part #: 2105617-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SINGLE, TERMINATED LASER DIODE AMP CONNECTOR, 105MM LONG, PHOTONIC FRU: Yes Status: Orderable Part #: 2105617-4 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: SINGLE, TERMINATED LASER DIODE AMP CONNECTOR, 105MM LONG FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY. COMMON TO MFG MS26561 DSM 1/7/01 PART OF 2229971
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required