Pular para o conteúdo principal
GE Healthcare · CT2000004242000004162000004152005200154294967295

Lightspeed 1.X - Octane

Symptom

System will not come up to applications with image space errors. System will not come up to applications with image space errors

Error message

Error: 9054 Cannot access directory in read/write mode : /usr/g/sdc_image_pool/buf_image. See LD for more details Error: 9054 Cannot access directory in read/write mode : /usr/g/sdc_image_pool/buf_image. Mon Feb 27 07:17:33 2006 Error: 9054 Host: csbipct2 Process: dbwserver File: init.c Line: 93 Cannot access directory in read/write mode : /usr/g/sdc_image_pool/buf_image

Error: 200000424 Image available space failed. See the long description for the full error Sat Jan 22 18:25:36 2005 Error: 200000424 Host: chr09bc1cti Process: dms File: AvailSpace.cxx Line: 155 Image available space failed. code: 200000416 Or Sat Oct 28 06:55:39 2006 Error: 200000424 Host: ct02 Process: dms File: AvailSpace.cxx Line: 155 Image available space failed. Error code: 200000415

Error: 200520015 Updating image space, received a bad status code: See the long description for different status codes Sat Jan 22 18:23:06 2005 Error: 200520015 Host: chr09bc1cti Process: environmentMgr File: ImageSpaceFormMgr.c Line: 117 Updating image space, received a bad status code: 200000416 Reason: Response from SdC timeout you may also see this error with different status codes, such as Fri Jul 28 05:12:48 2006 Error: 2

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2169940-81 Octane Computer 73 GB Disk Drive - replaces 2169940-78 (18GB) See long description for aditional information 2169940-81 Octane 73 GB Disk Drive, replaces 18 GB Disk 2169940-78 Part #: 2169940-81 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: CT OCTANE 2 73 GB DISK DRIVE FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.