LightSpeed 5.X (60kW)

Symptom

Grinding noise coming from the table while driving cradle in or out. Table may move slower than normal. Customer may hear a grinding noise coming from the table while driving cradle in or out. Table may move slower than normal.

Error message

Error: 260115236 MTCB Cradle Drive did not stop when expected Tue May 30 16:57:09 2006 Error: 260115236 Host: tgp Process: tgp_subsys File: mtcb_cradle_monitor.c Line: 487 Tue May 30 16:57:08 2006 Host : Table Sub-System Ermes # : 260115236 Exception Class : Soft Severity : Pri/Soft File : mtcb_cradle_monitor.c Line# : 487 Function : No System Function Reported Scan Type : None/Unknown Scan: 17521/3/0 MTCB Cradle Drive did not stop when expected. Target position = 4294967295 Current Position = 132453 Start Position = 143131 Expected end time = 1510 Current time = 1530 Target velocity = 0 Current velocity = -50450

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution