Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT18414217231184188170851713517140

LightSpeed 2.X (Plus, QX/i) - GRE

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 184142 ISD Missed See Long description for complete error Wed Nov 8 11:01:40 2006 Error: 184142 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1043 Function: No System Function Reported Scan: 496/2/2 Type: Axial Exception Level: Pri/Soft Time: 11:01:40:295 Log Series: 937 ISD missed. Rx ISD: 1000 mS Requested ISD = 1 Second Actual ISD: 17231 Actual ISD = 17 Seconds More Data Follows: See Error: 184188 for additional detail

Error: 184188 Timing (in mS) since end of last scan Col:17085 DvRdy:17135 GenX:17135 TrigEn:17140 ExpEn:17230 See Long Description for more details Wed Nov 8 11:01:40 2006 Error: 184188 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1048 Function: No System Function Reported Scan: 496/2/2 Type: Axial Exception Level: Pri/Soft Time: 11:01:40:295 Log Series: 938 Last Hdw Rdys- Offsets:0x40 Cpb:0x20 Xray:0x80 Timing (in mS) since end of last scan. RxRcvd:-670 DoRdy:95 OfSt:165 Cpb:17085 OfStp:17095 Cr:810 Ax:100 GenP:175 Das:160 Col:17085 DvRdy:17135 GenX:17135 TrigEn:17140 ExpEn:17230 (Rdy:0xf00e2fff Of:0x540 Auto:0x57c Cpb:0x4057c Xray:0x80bff) This Error code is can give you a clue as to where the problem is comin

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5140001-3 (replaces 5140001-2) Helios 3 collimator with notched filler bar and rating plate (used on 1.X-4.X) See Long Description for more info Part #: 5140001-3 Part Type: GP REPAIRABLES Part Weight: 115.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 3 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: ** FE Assist ** 1.x 2.x 3.x 4.x started with Ultras added minor quality improvements to the dash 2 and is backwards compatible (12/6/06 212001261) Part #: 5140001-2 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 2 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: This part is replacement for 2214768-2, please do not use stock of 2214768-2. ms13452 050516 Can use the 5140001-3 if no Stock of the -2 (9/26/05 212001261)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues