MPX L300 no 105 exposures.

Symptom

MPX L300 no 105 exposures. No 105 exposure, fluoro and spot films were ok. The problem was traced to no image gate movement during prep of the 105 camera. If the switch (152 on the mx1a14a1) that is used to swap positions of 105 and camera for a two port system was actuated, the image gate switched normally, the only difference was an inverter on that card. It was found to be bad. PHOTOSPOT