Ultranet SN 3299 Ultranet SA Collimator Setting the current feedback pots for Servo Damping

Symptom

What is the proper procedure to set the Current feedback pots for Servo Damping on an Ultranet SA Collimator To set up the current feedback pots P1 and P2 on PWA 45296175 for servo damping. The current revision of the system steering guides does not call out the servo damping adjustment. Since a particular collimator head and electronics are not tested together in the factory, the servo damping must be adjusted at the customer site during installation. The steering guides will be updated at their next revision.